Quick Answer: Which EC, CF or ppm Meter Should I Use for Hydroponics?
- EC, CF and ppm are the same conductivity reading shown three ways. CF is EC x 10, ppm 500 is EC x 500 and ppm 700 is EC x 700.
- Set your meter to the scale your nutrient label or feed chart uses, and never mix scales. A 700 scale chart read on a 500 scale pen means a feed about 40% too strong.
- A pen like the Bluelab Digital EC Meter suits most home growers. A Combo Meter reads pH and EC together, and a Guardian Monitor watches the tank all day.
- Buy a meter with automatic temperature compensation. Without it, a warm reservoir reads stronger than it is.
- Bluelab conductivity meters are factory calibrated. Clean the probe and check it in fresh 2.77 EC solution about once a month.
- Most wrong readings come from a dirty probe, a missing shroud, the wrong scale or an unmixed tank, not a broken meter.
Products for This Guide
- Bluelab Digital EC Meter, $159.95. EC pen that reads EC, CF and ppm, with automatic temperature compensation.
- Bluelab Combo Meter, $424.95. pH, EC and temperature in one reading. The EC probe needs no calibration.
- Bluelab Calibration Solution, $64.95. Box of 30 single use 2.77 EC sachets for checking your meter.
The best EC meter for hydroponics is one that reads in the same scale as your nutrient label, corrects for water temperature, and gets a clean and a check in 2.77 EC solution once a month. For most home growers that's a waterproof EC pen. Step up to a combo meter if you want pH and EC from one probe, or a continuous monitor if you can't check the tank every day.
EC vs CF vs ppm: Same Reading, Different Numbers
Every one of these meters measures one thing: how well your nutrient solution conducts electricity. More dissolved fertiliser salts means higher conductivity. The meter then shows that number on whichever scale you pick.
- EC (electrical conductivity) is the base reading, in millisiemens per centimetre (mS/cm). Bluelab recommends EC because it's the international standard.
- CF (conductivity factor) is EC x 10. CF 16 is EC 1.6. It's an older scale you still see on Australian nutrient labels, feed charts and meters.
- ppm 500 (also sold as TDS) is EC x 500. EC 1.6 reads 800 ppm.
- ppm 700 is EC x 700. EC 1.6 reads 1120 ppm.
A meter can't measure ppm directly. It measures conductivity and converts it, so ppm is only an estimate. That's also why there's no such thing as a separate ppm sensor: a ppm pen is an EC pen with a ppm setting.
| EC (mS/cm) | CF | ppm 500 | ppm 700 |
|---|---|---|---|
| 0.8 | 8 | 400 | 560 |
| 1.2 | 12 | 600 | 840 |
| 1.6 | 16 | 800 | 1120 |
| 2.0 | 20 | 1000 | 1400 |
| 2.4 | 24 | 1200 | 1680 |
For any other number, use our EC to ppm converter. It converts between EC, CF and all the common ppm scales.
Which ppm scale should I use?
Use the scale your nutrient label or feed chart is written in. Bluelab's own advice is that an Australian feed chart could be on any of the common ppm scales, so check before you mix. If the chart gives EC or CF as well, use that instead and skip ppm altogether.
Here's why it matters. Say a chart asks for 1400 ppm on the 700 scale, which is EC 2.0. If your pen is set to the 500 scale and you mix until it reads 1400, you've made EC 2.8. That's 40% stronger than the chart intended, and well above what lettuce or herbs want. Some meters also offer a third scale around 640 or 650, which adds another way to get it wrong.
The fix is to pick one scale, set it on your meter, and write it on the nutrient bottle. If you're not sure what EC your crop needs, our guide on what EC to run covers the targets.
Pens, Combo Meters and Continuous Monitors
| Type | What it does | Suits |
|---|---|---|
| EC pen | Handheld, dip it in the tank or a sample. EC and temperature only. | Most home growers with one reservoir. Pair it with a pH pen. |
| Heavy duty EC meter | Long body that reaches into deep tanks. Reads on contact with an LED scale. | Deep reservoirs and daily use where you want to stir and read in one go. |
| Combo meter | pH, EC and temperature from one unit and one reading. | Recirculating, DWC, NFT and coco feed programs where you check pH and EC together. |
| Continuous monitor | Probes stay in the reservoir and read all day, with alarms. | Growers who can't check daily, or where one bad day costs a whole crop. |
| Root zone meter | Probes go into the growing media, not the tank. | Coco and soil growers who want the EC the roots are sitting in. |
A pen tells you what the tank was doing when you checked it. A monitor tells you what it's doing now. Most growers start with a pen and only move to a monitor once they've been caught out by something that happened overnight.
Temperature Compensation
Conductivity rises as water warms. For typical salt solutions it climbs by about 1.5 to 2.2% for every degree Celsius. So a reservoir reading EC 2.0 at 20 degrees would read about 2.3 to 2.4 at 30 degrees if nothing corrected for it, even though nothing in the tank changed.
Automatic temperature compensation (ATC) fixes this. The meter reads the water temperature and converts the EC to what it would be at a reference temperature, usually 20 or 25 degrees. That's what lets you compare a reading from a hot January afternoon with one from a cold July morning.
- Check that any meter you buy says ATC or automatic temperature compensation. The Bluelab Digital EC Meter, Combo Meter and Pulse all list it.
- Give the probe time to reach the water temperature. Bluelab says errors can occur when the solution is hotter or cooler than the probe, and suggests leaving it in for 1 to 2 minutes. That matters when a meter comes out of a cold shed into a warm tank.
- ATC corrects the reading, not the water. If your reservoir is sitting above about 22 degrees in summer, that's a separate problem. See why your reservoir is too hot.
How to Calibrate an EC Meter
Calibration means telling the meter what a known solution should read. Which solution you use depends on the meter, so check the manual first.
- Bluelab meters use 2.77 EC standard solution. Many other brands use 1.413 mS/cm instead. Only calibrate in the solution your meter's manual names. A meter set up for 1.413 won't calibrate properly in 2.77.
- Use fresh solution every time. Bluelab says old or contaminated solution makes a meter read high. Single use sachets like the Bluelab 2.77 EC sachets avoid that. Don't pour used solution back or dip a probe into the bottle.
- Clean the probe first. Calibrating a dirty probe locks the error in.
Do Bluelab EC meters need calibrating?
Mostly not, and this is where EC meters differ from pH meters. Here's what Bluelab's documentation says for the meters we stock:
- Heavy Duty EC Meter (Truncheon style): factory calibrated, no calibration needed. Bluelab still says to clean it and test it in 2.77 EC solution at least once a month.
- Digital EC Meter (pen): comes factory calibrated. Bluelab says you don't need to calibrate it out of the box, but if it reads wrong in fresh 2.77 solution after a clean, calibrate it. Rinse, put it in fresh 2.77 solution, hold the calibration button for 3 seconds, and wait for the confirmation.
- Combo Meter and Guardian Monitor: the conductivity and temperature probe needs no calibration. Only the pH probe does.
- Bluelab pH EC Temperature Meter Bluetooth: pH takes a two point calibration. The conductivity side needs none.
For meters that don't calibrate, the 2.77 solution is a check, not a calibration. Clean the probe, dip it in fresh solution, and look at the reading. Bluelab's cleaning guide says it should come back within 0.1 EC of the standard. If it doesn't, clean it again. If it still won't read right after a proper clean and fresh solution, the probe may be damaged.
Cleaning the Probe
A film of nutrient, oil or grime on the probe face makes an EC meter read low. Bluelab recommends a clean every month. Their method for conductivity probes:
- Take the shroud off the probe. On the Truncheon style meter you twist it a quarter turn. On others, warm it in your hand and pull it off.
- Put one or two drops of conductivity probe cleaner on the probe face.
- Rub the face with a chamois.
- Rinse under running tap water while you keep rubbing, until every trace of cleaner is gone.
- Look at the water on the probe face. It should sit as an even film. If it beads up, clean it again.
- Put the shroud back on without touching the probe face, then check it in fresh 2.77 EC solution.
The Bluelab Probe Care Kit has the conductivity cleaner, chamois and 2.77 sachets in one box. Between cleans, rinse the probe in clean water after every reading. Unlike a pH probe, an EC probe doesn't need to be stored wet.
Why Is My EC Meter Reading Wrong?
Before you blame the meter, work through these in order. Most are quick to check.
- Wrong scale. The meter is on ppm 500 and your chart is ppm 700, or it's on CF and you're reading it as EC. Check the units on the screen.
- Dirty probe. Bluelab lists a dirty probe head as the usual cause of low readings. Nutrient residue and the oils from organic nutrients build up on it.
- Fingers on the probe. Skin oil on the probe face also pulls readings down. Hold the meter by the body.
- Shroud missing or loose. The shroud has to stay on when you read. Without it, readings are off.
- Probe not settled. Leave it in the solution for a minute or two so it matches the water temperature.
- Tank not mixed. Right after you add nutrient, the solution near where you poured is stronger. Let the pump or air stone run before you read, or stir a sample.
- Probe too close to the side or bottom. Keep it about 2.5 centimetres clear of the walls of the cup or tank.
- Old calibration solution. A reused or opened bottle of standard makes a good meter look bad. Use a fresh sachet.
- Reading plain water. RO water, rainwater and distilled water have almost no conductivity. The Truncheon style meter won't turn on below 0.2 EC, so it can't read them at all.
- Flat battery. If the display won't come on or keeps dropping out, replace the battery before you go further.
If the meter checks out fine in 2.77 solution, the reading is probably right and something is changing in the tank. Our guide on why EC keeps rising or falling covers the usual causes and what to do about each. Your starting water counts too. Hard tap water or bore water can add a fair bit of EC before any nutrient goes in, so check it on its own first. Our Australian tap water guide has more on that.
Testing Runoff vs the Reservoir
What you test depends on your system.
- Recirculating systems (DWC, NFT, ebb and flow): the solution comes back to the tank, so the reservoir is the number that matters. Test it at the same time each day with the pump running.
- Drain to waste in coco or rockwool: test what goes in and what comes out. Measure your feed in the tank, then catch the runoff from the same watering and test that.
Runoff EC is a rough guide to what's building up in the media. Bluelab says runoff gives a good approximation of available nutrient, suggests about 20% runoff, and says runoff EC much higher than your feed EC means salts are building up or the plants aren't taking up enough. A common commercial guideline is to keep runoff no more than 30% above your feed. So with a feed of EC 2.0, runoff above about 2.6 is a sign to water through with plain pH adjusted water until runoff comes back down to your feed EC.
If you grow in coco or soil and want the root zone reading without collecting runoff, the Bluelab Pulse Meter reads EC, moisture and temperature in the media itself. It needs the media above 20% moisture for an accurate EC reading, and it doesn't suit pure perlite or rockwool. Our coco coir guide has more on feeding in coco.
What to Buy at Each Budget
- Around $160: an EC pen. The Bluelab Digital EC Meter at $159.95 reads EC, CF and ppm, has ATC and is waterproof. Pair it with a pH pen. This is the right first meter for one reservoir.
- Around $195: no calibration, deep tanks. The Bluelab Heavy Duty EC Meter at $194.95 reads EC, CF, ppm 500 and ppm 700 on an LED scale, with a long body for deep reservoirs. Its scale runs up to EC 3.6.
- Around $385 to $425: pH and EC in one. The Bluelab pH EC Temperature Meter Bluetooth at $384.95 logs readings to your phone. The Bluelab Combo Meter at $424.95 has a 2 metre cable and a replaceable pH probe. Either replaces a pH pen and an EC pen.
- Around $475: coco and soil. The Bluelab Pulse Meter at $474.95 reads EC in the root zone instead of the runoff. It doesn't read pH.
- Around $525: continuous monitoring. The Bluelab Guardian Monitor WiFi at $524.95 reads pH, EC and temperature all day and sends alerts to your phone through the Edenic app.
- Running costs. A box of 30 2.77 EC sachets is $64.95, and the Probe Care Kit is $44.95. At one check a month, a box of sachets lasts more than two years on one meter.
If you also need a pH meter, our pH meter guide covers pens, probes and buffers.
Frequently Asked Questions
What's the difference between EC, CF and ppm?
They're the same conductivity reading on different scales. CF is EC x 10, ppm 500 is EC x 500 and ppm 700 is EC x 700. So EC 1.6 is CF 16, 800 ppm on the 500 scale and 1120 ppm on the 700 scale.
Which ppm scale should I use?
Use the scale your nutrient label or feed chart is written in, and set your meter to match. If the chart also gives EC or CF, use that instead. Mixing scales is the most common way growers end up with a feed far stronger or weaker than they meant.
How do I calibrate an EC meter?
Clean the probe, rinse it, then put it in fresh calibration solution and follow the meter's calibration steps. Bluelab meters use 2.77 EC solution, while many other brands use 1.413 mS/cm. Only use the solution your meter's manual names, and use a fresh sachet each time.
Do Bluelab EC meters need calibrating?
Bluelab conductivity meters are factory calibrated. The Truncheon style meter, the Combo Meter's EC probe and the Guardian Monitor's EC probe need no calibration. The Conductivity Pen can be calibrated in 2.77 EC solution if it still reads wrong after cleaning. Bluelab recommends cleaning and checking all of them in 2.77 EC solution about once a month.
Why is my EC meter reading wrong?
The usual causes are the wrong scale on the display, a dirty probe, skin oil on the probe face, a missing shroud, a probe that hasn't settled to the water temperature, or a tank that hasn't mixed. Clean the probe and check it in fresh 2.77 EC solution. If it reads right there, the meter is fine and the change is in your tank.
Should I test runoff or the reservoir?
In a recirculating system, test the reservoir. In drain to waste coco or rockwool, test both the feed and the runoff from the same watering. Runoff well above your feed EC means salts are building up in the media.
What's the best EC meter for hydroponics?
For most home growers with one reservoir, a waterproof EC pen with automatic temperature compensation is enough. A combo meter makes sense if you want pH and EC from one probe, and a continuous monitor if you can't check the tank every day.